MULTICHARME: A modified Chernin-type multi-pass cell designed for IR and THz long-path absorption measurements in the CHARME atmospheric simulation chamber

Éric Fertein
Nicolas Houzel
Guillaume Dhont
Robin Bocquet
  • Fonction : Auteur
  • PersonId : 1196820
  • IdRef : 096652578
Gaël Mouret

Résumé

Abstract. We have developed MULTICHARME, a modified Chernin-type multi-pass cell especially designed for IR and THz long-path absorption measurements in the CHamber for Atmospheric Reactivity and Metrology of the Environment (CHARME). By measuring the output power using a near-IR diode-laser and a THz amplified multiplication chain, we have established that the effective reflectivity of MULTICHARME is better than 94 % over approximately three decades of frequency. Absorption measurements of N2O have been performed by probing highly excited rovibrational transitions in the near-IR and ground state rotational transitions at submillimetre wavelengths. In each case the linearity of the absorbance with the pathlengths was verified. Finally, we demonstrate that THz spectroscopy is able to study the isotopic composition of greenhouse polar gases such as N2O and to absolutely quantify stable (N2O) and reactive (O3) species at trace levels. Moreover, a THz monitoring at low pressure of the ozone decay in the chamber has been performed. The deduced ozone lifetime of 3.4 ± 0.1 h is shorter compared with previous measurements performed in CHARME at atmospheric pressure. For the first time, the ability of THz rotational spectroscopy to monitor, with a very high degree of selectivity, stable and reactive polar compounds at trace level in an atmospheric simulation chamber is demonstrated. However, the sensitivity of the THz monitoring needs to be improved to reach the atmospheric trace levels. For this purpose, it is necessary to figure out the baseline variations as well as possible induced by the multiple standing waves present in MULTICHARME.
Fichier principal
Vignette du fichier
amt-15-1201-2022.pdf (4.24 Mo) Télécharger le fichier
Origine : Fichiers éditeurs autorisés sur une archive ouverte
Licence : CC BY - Paternité

Dates et versions

hal-04097350 , version 1 (16-11-2023)

Licence

Paternité

Identifiants

Citer

Jean Decker, Éric Fertein, Jonas Bruckhuisen, Nicolas Houzel, Pierre Kulinski, et al.. MULTICHARME: A modified Chernin-type multi-pass cell designed for IR and THz long-path absorption measurements in the CHARME atmospheric simulation chamber. Atmospheric Measurement Techniques Discussions, 2022, 15 (5), pp.1201-1215. ⟨10.5194/amt-2021-399⟩. ⟨hal-04097350⟩
37 Consultations
7 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More